Care, maintenance and trouble-shooting
Axiostar plus Troubleshooting Carl Zeiss
B 40-815 e 12/01 4-3
4.2 Troubleshooting
Troubleshooting for the Axiostar plus is
described in further detail using the following
two examples:
changing the fuses and
changing the 6 V, 20 W halogen lamp
Further measures are summarized in the table
under4.2.2.
4.2.1 Changing the fuses
Switch off the instrument via the on/off
switch (4-1/1), disconnect line plug from
the line and disconnect line connector from
instrument socket.
Check line cable and plugs and exchange
them, if required.
Use a 2.0 mm clockmakers screwdriver to
press in the two spring shackles on the
sides of the fuse holder over the slot and
remove the fuse holder (4-1/3).
Remove defective fuse inserts (4-1/2) from
the fuse holder and replace them with new
ones
Reinsert the fuse holder containing the new
fuses; the spring shackles on the sides must
audibly lock into place.
T 0.8 A; 250 V; 5 × 20 mm fuse inserts must be
used.
For the catalog numbers of fuse inserts, please
see Section 4.2.4.
1 On/off switch
2 Fuse inserts
3 Fuse holder
Fig. 4-1 Changing the fuses