Operation
Carl Zeiss Attaching photomicrography equipment Axiostar plus
3-18 B 40-81 e 12/01
☞
For the use of the photomicrography equip-
ment, please observe the relevant separate
manuals in addition to the information provided
in this manual.
1 Cable release
2 Camera housing
3 T2-Adapter for CONTAX bayonet
42.5× connector for T2
5 Phototube
6 Hexagonal screw
7 Control
8 Dust protection cap
Fig. 3-13 Attaching SLR camera,
e.g. CONTAX ARIA
3.9 Attaching photomicrography
equipment
The Axiostar plus microscope with phototube
can be changed from observation to photo-
micrography via the control knob (3-13/7 and
3-14/8) attached to both sides of the
phototube (photomicrography position: turned
counterclockwise until stop: 100 % doc). In the
other position of the knob, the beam path is
directed to the eyepieces (100 % vis). Special
adapters allow commercially available 35 mm
SLR cameras and special microscope cameras
(e.g. MC 80
DX) to be attached to the camera
port of the Axiostar plus.
If focusing is not to be made via the viewfinder
of the camera, the component with the
eyepiece reticle must be screwed in the
eyepieces (see sections 1.6 and 2.5.1).
• For focusing, set the eyepiece until the
reticle is imaged in focus.
• Then focus until both the microscope image
and the reticle are in focus.
3.9.1 Attaching SLR camera,
e.g. CONTAX ARIA
• Screw T-2 adapter (3-13/3) for the CONTAX
bayonet on the 2.5x connector for T2
(3-13/4) (456005-0000-000).
• Attach camera housing (3-13/2) and cable
release (3-13/1), if required.
• Loosen three hexagonal screws (3-13/6)
remove the dust cover (3-13/8) from the
phototube (3-13/5) and insert the
premounted unit (3-13/A) in the phototube.
• Align the camera unit in the required
position and tighten the three hexagonal
screws (3-13/6).
• For photomicrography, turn control (3-13/7)
counterclockwise until stop. When using
the phototube 30°
ICS, pull out the lever for
light path switching completely (100 %
light to camera port).