Care, maintenance and trouble-shooting
Carl Zeiss Care and maintenance of the instrument Axiostar plus
4-2 B 40-815 e 12/01
4 CARE, MAINTENANCE AND
TROUBLE-SHOOTING
4.1 Care and maintenance of the
instrument
Care of the Axiostar plus is limited to the
following operations:
Cover the instrument with the dust cover
after every use.
Do not set up the instrument in a damp
room, i.e. max. humidity < 85 %.
Cover open tubes with dust protection
caps.
Remove dust and dirt from visible optical
surfaces using a brush, airblower, Q-tip,
optics cleaning paper or a cotton cloth.
Remove water-soluble dirt (coffee, Coke,
etc.) after breathing on it and wiping it off
with a dust-free cotton cloth or a
moistened cloth. A mild cleaning agent can
also be added to the water.
Remove stubborn, oily or greasy dirt
(immersion oils, fingerprints) with a wad of
cotton or a dust-free cotton cloth dipped in
the optics cleaning mixture L.
This cleaning mixture is produced of
90 Vol% benzoline and 10 Vol%
isopropanol (IPA). The various components
are also known under the following
synonyms:
benzoline: medical alcohol,
petroleum ether
Isopropanol 2-propanol,
dimethyl carbinol,
2-hydroxypropane
Clean the optical surface by moving in circles
starting in the middle. Slight pressure should be
exerted on the optics during cleaning.
When using the Axiostar plus in humid climatic
zones, proceed as follows:
Store the Axiostar plus in bright, dry and
well-ventilated rooms with a humidity of
less than 85 %; store particularly sensitive
components and accessories, such as
objectives and eyepieces, in a dry closet.
The risk of growth of fungus on opto-
mechanical instruments always exists in the
following conditions:
Relative humidity of more than 75% and
temperatures between +15 °C and +35 °C
for more than three days.
Installation in dark rooms without air
ventilation.
Dust deposits and fingerprints on optical
surfaces.