Operation
Axiostar plus Attaching
ICS double tube Carl Zeiss
B 40-815 e 12/01 3-17
3.8 Attaching
ICS double tube
• Loosen knurled screw (3-12/4) on the stand
and remove binocular tube, if available.
• Remove dust cover from the underside of
the tube.
• Press the
ICS double tube (3-12/1) to the
right in a slightly inclined position into the
spring of the stand mount.
• Then set the double tube
ICS by lowering
the left side into the dovetail of the stand
mound. This will secure the double tube
ICS
from falling out.
• Turn the
ICS double tube into the required
viewing direction and use the knurled screw
(3-12/4 or hexagonal screw) to clamp it into
the threaded opening (3-12/5). Both the
illustrated rotation and the 90° backward
rotation are possible. Accordingly, the
controls of the
ICS double tube are
accessible to the user from the side or from
the front.
• Remove dust caps from the eyepiece tubes
and insert two eyepieces each (3-12/2) with
the same field number 18 or 20.
• Insert country-specific adapter (3-12/9) into
the power unit (3-12/8).
• Insert angled connector (3-12/10) of the
power unit into the 12 V 3-12/3) of the ICS
double tube.
• Use the enclosed self-adhesive cable holders
(3-12/11) to suitably attach the cable to the
stand and guide it to the back.
• Connect power unit to the line.
The ICS double tube is operated via:
• Adjusting lever (3-12/7) for the positioning
of the light pointer in the microscope image
• Control (3-12/6) to set the brightness of the
light pointer.
1 ICS double tube
2 Eyepieces
3 12 V socket
4 Knurled screw (or hexagonal screw)
5 Threaded opening for knurled screw or hexagonal
screw
6 Control for intensity of light pointer
7 Adjusting lever for light pointer
8 Power unit
9 Country-specific adapter
10 Angled connector
11 Cable holder
Fig. 3-12 ICS double tube