Carl Zeiss Attaching epi-fluorescence illuminator Axiostar plus
2-26 B 40-815 e 12/01
• Hold the new burner on the labeled metal
base and place it in the heat sink so that
the reflective portion of the burner points
down when placed in the lamp frame. If
both sides are reflective, make sure that the
electrode labeled “UP“ is at the top. If there
is a sideways reflective melting position on
the burner, it must point to the side in
relation to the beam path.
• Carefully tighten clamping screw on the
heat sink.
Avoid getting fingerprints on the glass
components of the burner. If fingerprints do
occur, remove immediately.
• Press down spring handle (2-23/4) and
place burner with heat sink into the lamp
frame (2-23/5). Make sure to touch only the
heat sink.
• Slowly let go of spring handle and release
heat sink.
The heat sink must be parallel to the lamp base.
To aling, press spring handle and turn heat sink
with burner in the lamp frame.
• Replace lamp housing and tighten clamping
screws (2-23/1). Make a note of the
number on the operational hours counter.
Change the burner after reaching the expected
life of 100 hours.