Operation
Carl Zeiss Transmitted-light brightfield Axiostar plus
3-4 B 40-81 e 12/01
3.2 Using the transmitted-light
brightfield technique
3.2.1 General principle
Transmitted-light brightfield microscopy is the
most usual of all the optical microscopy
techniques, since it allows the easy and fast
viewing of high-contrast or stained specimens
(e.g. blood smears).
In order to obtain optimum resolution with full
illumination of the field, the condenser, the
luminous-field diaphragm and the aperture
diaphragm must be set in accordance with the
rules of the KÖHLER illumination principle.
Here, the illumination cone is adapted to the
objective's opening cone. In this way, the
numerical aperture of the optical system is
used, and "superfluous" light, which can cause
interference in the form of scattered light, is
prevented.
1 Brightness control
2 Aperture diaphragm lever
3 Condenser drive
Fig. 3-3 Transmitted-light brightfield for KÖHLER
illumination, preparation
3.2.2 Transmitted-light brightfield
configuration
Each Axiostar plus microscope can be
configured to permit the transmitted-light
brightfield technique.
When using the Abbe condenser with 5-
position turret, set the brightfield position
(H) via the turret disk.
If a slider for phase contrast is available with
the 0.9/1.25 Abbe condenser, it must be
removed when using the transmitted-light
brightfield technique.
3.2.3 Setting transmitted-light
brightfield for KÖHLER
illumination
Requirement:
As described in section 2, the microscope is
ready for operation and switched on
according to section 3.1.
Settings:
First, place a high-contrast specimen with
0.17 mm cover slip on top of the
mechanical stage 75x30 R/L.
Set image brightness using the brightness
control (3-3/1) on the microscope stand.
Move the Abbe condenser 0.9/1.25 to the
upper stop position via the condenser drive
(3-3/3) and move aperture diaphragm lever
(3-3/2) to the center position.
If the microscope is equipped with a mechanical
stage 75 x 30 R (stage drive on the right side),
then the condenser drive will be attached on
the left in the factory.
The condenser drive will be located on the right
side of the microscope if the microscope is
equipped with a mechanical stage 75 x 30 L.