MIU User Guide For Zeiss AxioVert
200 Inverted Fluorescence
Microscope
Molecular Imaging Unit (MIU)
University of Helsinki
www.miu.helsinki.fi
2.9.2009
Zobrazit stránku 0
1 2 3 4 5 6 ... 19 20

Shrnutí obsahu

Strany 1 - Microscope

1 MIU User Guide For Zeiss AxioVert 200 Inverted Fluorescence Microscope Molecular Imaging Unit (MIU) University of Helsinki www.miu.helsink

Strany 2

10 As a last step, adjust the iris diaphragm using the slider above the DIC condenser (Fig. 5). First, fully open it and then, if needed, close it as

Strany 3 - 1. General

11 Focusing is done by using the focusing knobs on either side of the microscope. The refresh rate of live image can be adjusted (Fig. 14). Quality o

Strany 4 - 2. Microscope use

12 Figure 13 Camera settings in AxioVision software In the image window there is a "French flag" button in the tool bar at the b

Strany 5

13 The recommended minimum camera resolutions for each objective are: Objective NA Recommended min. resolution 10x 0.3 3900 x 3090 20x 0.4 2600

Strany 6 - 2.3 Fluorescence filters

14 BestFit shows the whole range excluding specified percentage of actual minimum and actual maximum values Min/Max shows the whole range from actual

Strany 7

15 When you click with the right mouse button and select VIEW ANNOTATIONS, you will remove the scale bar. If you select VIEW ANNOTATIONS again, the s

Strany 8

16 You can save images temporarily to that folder and later move them into your own computer. The folder will be emptied occasionally (it may happen

Strany 9

17 taken!). There is a 3 GB limit for each user. It is recommended to transfer images somewhere else after your imaging session.

Strany 10 - 3. Imaging

18 5. Ending your imaging session When you have finished your imaging session you should: - Close the fluorescence light shutter (Fig. 3) - Log off

Strany 11 - Figure 12

19 The whole image is too bright: Is the transmitted light (halogen bulb) off (p. 5)? Cannot see the whole field of view: Is the fluorescence filter

Strany 12

2 CONTENTS 1. General ... 3 1.1 Instrum

Strany 13 - 3.3 Display properties

20 6.3 Computer It’s not possible to log on to the computer: Use HYAD user name. Check that the domain selected is ATKK. I do not have HYAD-username:

Strany 14 - Figure 15 Display tab

3 1. General 1.1 Instrument This user guide covers the use of the Zeiss AxioVert 200 inverted epifluorescence microscope. For more information on eac

Strany 15

4 1.5 Users’ mailing list All registered users are automatically added to the users’ mailing list. Please read those e-mails to know what is going on

Strany 16 - 4.8 Saving images locally

5 Figure 1 Adjustment of the light path to a correct port Arrow 1: ocular/camera selection Arrow 2: selection of the front/side port. Red circle:

Strany 17

6 2.2 Objectives There are following objectives in the microscope: Position Type Magn. NA Oil/Dry Phase 1 Plan-Neofluar 10x 0.3 Dry Ph1 2

Strany 18 - 6. Troubleshooting

7 2.4 Transmitted (bright-field) light microscopy Transmitted light microscopy is used for specimens that are stained, or have natural pigment so the

Strany 19 - 6.2 Camera

8 - Select a 10x objective and DIC .3-.4 condenser position by rotating manually the condenser wheel - Focus on the specimen - Close luminous

Strany 20 - 6.4 AxioVision software

9 Pull the analyzer located on the left side of the microscope to the left so that it is in position shown in (Fig. 10). Figure 7 The polarizer fo