Start-Up
Carl Zeiss Attaching condensers Axiostar plus
2-16 B 40-815 e 12/01
1 Condenser carrier
2 Centering screw
3 LED illuminator
4 Condenser drive
Fig. 2-15 Attaching LED illuminator
Be careful of light-emitting diode radiation
when using the LED illuminator. Do not look
directly into the illuminator with your bare eyes
from a short distance. When adjustments are
required, the intensity must be reduced
accordingly via the potentiometer, or a suitable
attenuator must be used.
The operation life with battery operation
strongly depends on the brightness set and the
loading capacity of the battery.
2.8.4 Attaching LED illuminator
Insert LED illuminator (2-15/3) in the
condenser carrier (2-15/1) in the same way
as the Abbe condensers.
Tighten both centering screws (2-15/2) until
the LED illuminator lies approximately in the
center of the beam path.
Use condenser drive (2-15/4) to move the
condenser carrier into the topmost position.
Power to the LED illuminator can be supplied
either via the line power unit or, independent
of the line, via a commercially available 9 V
Alkaline battery (9 V / Ni Cd battery block) in
the integrated battery compartment, or via an
external voltage source ranging from 11 to
14 V DC (e.g. 12 V car battery or battery
pack C). The external voltage source must
permit a constant current of up to approx.
120 mA to be loaded.
The LED illuminator provides pleasant, color-
neutral white light with a constant color
temperature (approx. 7500 K) independent of
the set brightness. It is suitable for objectives
ranging from 2.5x to 100x. The light-emitting
diodes have a long life and cause only minor
operating costs.
Operation with power unit
Insert country-specific adapter (2-16/10)
into the power unit (2-16/9).
Insert angled connector of the power unit
(2-16/9) into the +12 V socket (2-16/2) of
the LED illuminator (2-16/1).
Connect power unit to the line.
Attach the cable to the back of the stand
with the self-sticking cable ties included.