Start-Up
Carl Zeiss Attaching condensers Axiostar plus
2-18 B 40-815 e 12/01
1 Condensor carrier
2 Spring pin
3 Centering screws
4 Groove
5 Darkfield condenser frame
6 Darkfield condenser
7 Fastening ring
Fig. 2-17 Attaching darkfield condenser
2.8.5 Attaching darkfield condenser
The darkfield condenser frame (for Axiostar
plus) is required to place darkfield condensers in
the condenser carrier.
• Place darkfield condenser (2-17/6) in the
darkfield condenser frame (2-17/5) and
screw on fastening ring (2-17/7)
☞
The dry darkfield condenser (0.7/0.85, 465506-
000-000) is screwed directly into the darkfield
condenser frame without the fastening ring.
• Press the darkfield condenser frame
(2-17/5) against the spring pin (2-17/2) into
the condenser carrier (2-17/1).
• Tighten both centering screws (2-17/3) on
the condenser carrier (2-17/1) until they
grip the ring mount and hold the darkfield
condenser frame (2-17/5) in place.