Axiovert 25, Axiovert 25 C, Axiovert 25 CFL
B 40Ć012 e 08/99 2Ć15
NOTE Shifting the VAREL illumination to outside the pupil corresponds to oneĆsided
darkfield illumination.
Shifting the VAREL illumination between the Ph and VAREL rings of the objective
corresponds to inclined brightfield illumination.
The middle engaging position of the Phā/āHā/āVar slider permits brightfield
illumination.
The left engaging position of the Phā/āHā/āVar slider permits the fast change from
VAREL to Ph contrast.
Figure 2Ć20 Pupil images in VAREL contrast
Objective
VAREL contrast
Inclined brightfield
illumination
OneĆsided darkfield
illumination
Ph ring VAREL ring VAREL diaphragm
Preparing the Var/H/Var slider for phase contrast
Remove the attenuation filter (2Ć21ā/ā3) from
the center position of the slider (remove
spring and filter).
Insert the ring diaphragm mount (2Ć21ā/ā1)
(contained in the delivery package of the
slider).
If required, loosen centering screws (2Ć21ā/ā2)
using the key.
Insert ring diaphragm in ring diaphragm
mount and center the ring diaphragm with
regard to the phase ring of the objective
using the centering screws (one centering
screw at each side of the slider).
Figure 2Ć21Varā/āHā/āVar slider
1
2
3