Axiovert 25, Axiovert 25 C, Axiovert 25 CFL
08/99 B 40Ć012 e
2Ć6
(7) Inserting the eyepieces
Remove the dust covers (2Ć8/1) and insert the
fixed eyepieces (2Ć8/2, 3) and the focusing
eyepieces in the eyepiece tube (2Ć8/4).
NOTE Depending on the configuration,
one or two eyepieces may permit
focusing to compensate for
differing degrees of myopia of
the two eyes.
(8) Compensating for myopia
1) with one fixed and one focusing eyepiece
Focus on the specimen via the fixed eyepiece (2Ć8/2) and the focusing drive. Focus on the
specimen for the second eye via the eyelens of the second eyepiece (2Ć8/3).
2) with two focusing eyepieces
Use the eyelens of the focusing eyepiece (2Ć8/3) to focus on the line figure of the eyepiece reticle
or, if there is no eyepiece reticle, on the edge of the field of view.
Focus on the microscope image of a specimen via the focusing drive by looking through the
eyepiece with reticle.
When the image and the eyepiece reticle are in focus in the above eyepiece, focus the image for
the second eye via the eyelens of the second eyepiece.
NOTE The position of the focusing drive on the stand must not be changed.
Figure 2Ć8 Inserting the eyepieces