Axiovert 25, Axiovert 25 C, Axiovert 25 CFL
B 40Ć012 e 08/99 2Ć7
(9) Using the binocular tube
The eyepiece distance is matched to the indiĆ
vidual interpupillary distance by swinging
the eyepiece tubes symmetrically towards
one another.
A higher (2Ć9ā/āA) or lower (2Ć9ā/āB) viewing
height is achieved by swiveling the tube.
(10) Inserting an eyepiece reticle (dia. 26 mm)
The WĆPl eyepieces 10x/18 foc. are intended
for use with reticles.
The image shift produced by the additional
path through the glass is taken into account
on the diopter scale by the fact that the zero
position is indicated not by the white dot
W
but by the red dot
R
.
The reticles (2Ć10/2) have been adhered to
screwĆin mounts (2Ć10/1) to allow easy
replacement.
To replace a reticle, unscrew the entire mount
and replace it with one containing the
required reticle.
Finally, check the image focus and readjust it,
if required (see section 2.1.2 (8), page 2Ć6).
Figure 2Ć9 Positions of the binocular tube
B
A
Figure 2Ć10 Inserting an eyepiece reticle