Multidimensional setup
Multiple channels
open "Acquisition Setup" dialogue
tick "Change channels using light paths"
add as many channels as you need with
the "+" button
from the appearing drop down menus
choose the desired channel presets
if you don`t want to do a stack, time series
or multiple positions set timepoint to 1,
choose no focus and no xy
Z stacks
Before any other adjustments:
Click on Z position button and set "Top" and "Bottom" to the mechanical limits
using defined focus positions
move focus to mid focus plane
click on Z position button, choose "Set Zero"
move focus to top position
click on Z position button, choose "Set Top"
move focus to bottom position
click on Z position button, choose "Set Bottom"
In “acquisition setup” “Channels/Z” tab, set the “Change focus using" to "Zeiss
Focus Drive"
Choose and set Z-spacing or number of slices
If using Autofocus
set "Change focus using" to "Zeiss Focus Drive"
choose and set Z-spacing or number of slices
set "Auto-focus using" to "Zeiss Focus Drive"
choose the Channel for Autofocus