6 Maintenance
For general information on care and mainte-
nance, please refer to the operating instructions
G 42-311, B 40-016 and B 40-042.
Kindly also observe the following::
The universal rotary stage is a precision instru-
ment and must therefore be protected against
mechanical damage to ensure that settings
and adjustments are not lost.
Treat the segments with care (risk of scratch-
ing).
Glycerin is recommended as an immersion
agent due to it being water soluble.
Always clean the universal rotary stage imme-
diately on completion of your work. For this,
loosen the knurled screws of the upper seg-
ment and lift it out of the center part of the uni-
versal rotary stage together with the sample
and the lower segment.
Carefully separate these components sideways
and rinse off the immersion fluid with clean wa-
ter, or dab it off using wet cotton wool. Distilled
water is recommended for glycerin and light
gasolin for immersion oil. Use an optical clean-
ing cloth or leather for the actual cleaning
process.
The mechanical parts are cleaned in the same
way.
Caution!
The annular groove for the lower seg-
ment and the guideway for the Schmidt rulers
must be cleaned very thoroughly due to the
high-precision fit to eliminate the possibility of
functional defects.
The smoothness of the vertical objective ad-
justment must be checked at regular intervals.
On completion of your work, store the universal
rotary stage and all accessories in the storage
container supplied.